Detail PROTEK 6510
Protek 6510 100MHz Oscilloscope
Features:
Four traces may be simultaneously displayed in ALT-sweep
Five vertical display modes: CH1, CH2, Dual, Add, Subtract
Vertical mode triggering for displaying 2 unrelated signals
Trigger hold-off circuit for synchronizing complex waveforms
Sweep speeds to 2nS/ Div
12kV acceleration voltage for bright displays of low duty cycle signals
Z axis input
TV Synchronization for either line or field patterns
20MHz bandwidth limit switch for triggering signals in a noisy environment
Specifications
Vertical Amplifier ( CH 1 and CH 2)
Sensitivity 5 mV/ Div to 5 V/ Div in 7 calibrated steps, ( X5 Mag 1mV/ Div to 1 V/ Div)
Accuracy ± 3% ( ± 5% in X5 Mag)
Bandwidth ( -3dB) DC to 100MHz
Rise Time £ 3.5 ns, X5 Mag: 20MHz Bandwidth: £ 17 ns
Input Impedance 1MW + 25 pF
Input Coupling AC-Gnd-DC
Invert Channel CH2
Display Modes CH1, CH2, Add, Subtract, Dual ( Alt/ Chop)
Maximum Input Voltage 400V ( DC or AC peak)
CH1 Signal Output 20mV into 50W
Horizontal Sweep
Timebase A - Sweep Time 50ns/ Div to 0.2 s/ Div in 21 calibrated steps ( 1-2-5)
Timebase A - X10 Mag 5ns/ Div to 20ms/ Div
Timebase A - Accuracy ± 3% ( ± 5% in X10 Mag)
Timebase A - Hold Off Continuously adjustable to > 2X sweep time
Timebase B - Sweep Time 50ns/ Div to 10µ s/ Div in 8 steps ( 1-2-5)
Timebase B - Accuracy ± 3%
Timebase B - Delay Time Continuously variable from 0.2µ s to 2s
Timebase B - Delay Modes Continuous or triggered
Timebase B - Sweep Displays A, B, A alt B, B triggered, X-Y
X-Y Mode
CH1-X axis, CH2-Y axis Sensitivity Same as CH1 vertical amplifier
Bandwidth ( X-axis) DC to 2MHz
Phase Error £ 3° DC to 10kHz
Triggering
Trigger Modes Auto, Normal, TV-V, TV-L
Source Int. CH2, line ext.
Coupling AC
Slope Positive or negative
Trigger Sensitivity Int ( DC to 10MHz) : 0.5 Div
Int ( 10MHz to 100MHz) : 1.5 Div
Ext ( DC to 10MHz) : 200mV
Ext ( 10MHz to 100MHz) : 500mV
TV ( Int) : 1 Div Min
TV ( Ext) : 1V P-P
External Trigger Input Impedance: 1MW + 20 pF
Maximum Input Volts: 400V DC or AC peak
Z Axis Modulation
Sensitivity ± 5V P-P Intensity increases with a negative going voltage input
Bandwidth ( -3dB) DC to 2MHz
Input Impedance Approx. 2 KW
Calibrator Signal
Voltage 0.5V ± 3% P-P 1kHz Square Wave
Duty Cycle Approx. 50%
CRT
Type 6 " rectangular CRT with internal graticule. ( 1 Div = 1cm)
Acceleration Voltage 12 kV
General Specification
Power Requirements AC Input Voltage: 100V, 120V, 220V, 240V selectable
Frequency 50/ 60Hz
Power Consumption 55W
Operating Temperature 0 to 40° C at 35% to 85% relative humidity
DIMensions 5.25 " H × 12.67 " W × 14.5 " D
Weight 24 lbs.
Supplied Accessories Manual, 1 AC Power cord, ( 2) X10 Probes